Maintenance
Combine calibration services and annual maintenance to keep your equipment in top condition.
Accuracy
Precision and reliability of our services are ensured by a team of experienced technicians.
In-House or On-Site
Choose between convenient in-house or on-site calibration services.
Efficiency
Benefit from fast turnaround times to minimize downtime.
We perform ISO-17025 compliant calibrations of gas dividers, CFO kits as well as venturis and mass flow meters (MFC's). In addition to gas flow, our labs are also accredited for temperature and pressure sensors.
Our services range from a pure inspection to a so-called all-inclusive service, which covers necessary repairs in addition to calibration and maintenance. If the measured values are outside of the specified tolerances, an appropriate adjustment can be made.
A standard or ISO 17025 calibration certificate certifies that the calibrated device is working properly within the specified tolerances.
Calibrated Item |
Measuring Range |
Uncertainty |
Rel. Pressure Sensor |
- 1,000 hPa ... 2,200 hPa |
0.64 hPa |
Abs. Pressure Sensor |
30 hPa ... 700 hPa |
0.60 hPa |
> 700 hPa ... 1,300 hPa | 0.20 hPa | |
> 1,300 hPa ... 3,500 hPa | 0.60 hPa | |
Volume Flow MFC | 1 l/min ... 200 l/min | 0.1 l/min + 0.4 % of the mean value |
Volume Flow Venturi | 1 m3/min ... 145 m3/min |
0.42 % of the mean value (∆T ≤ 2 K during calibration) 0.62 % of the mean value (∆T ≤ 5 K during calibration) |
Volume Flow Rate of Nitrogen (N2) / Gas Divider SL |
0.0025 l/min … 10 l/min | 0.35 % |
Volume Flow Rate of Nitrogen (N2) / Gas Divider i60 |
0.01 l/min … 10 l/min | 0.38 % |
Volume Flow Rate of Propane (C3H8) / Critital Flow Venturi (CFO) Devices |
0.02 l/min … 10 l/min | 0.58 % |
Temperature Sensor | 5 °C ... 70 °C | 0.54 K |
DIN EN ISO/IEC 17025:2018
AVL Analytical Technologies received accreditation of its calibration laboratories in Neuss and Gaggenau according to DIN EN ISO/IEC 17025:2018 for gas flow measurements as well as temperature and pressure sensors by the DAkkS (Deutsche Akkreditierungsstelle GmbH).
Flexible Scope: Overview of Current Revisions in the Permanent Laboratory and On-Site:
Calibration Item | Measurement Conditions / Procedure |
Rel. Pressure | DKD-R 6-1:2014 Rev3_2020 |
Abs. Pressure | DKD-R 6-1:2014 Rev3_2020 |
Temperature Quantities (direct reading thermometers with resistance sensor) | DKD-R 5-1-2023 |